LP604350 Injector
The Akreos MI60 lens is inserted using the LP604350 single-use injector.
|
|

HPMC placement
The loading chamber is coated with an HPMC viscoelastic: a small amount in the opening of the cartridge tunnel, two lines in the lateral grooves (1).
|

Lens removal
The lens is removed from its holder by grasping the full optic with the forceps. The lens must be positioned at the top of the holder, anterior face up (2).
|

Lens loading
The lens is placed in the centre of the loading chamber and the haptics are pushed down inside the lateral grooves(4).
|

Cartridge loading
The cartridge is closed and locked after having checked that no lens material is trapped between the wings(5).
|

Plunger advancement
The cartridge is placed in the injector and the plunger is advanced slowly until the silicone cushion has reached the tunnel opening. The lens is ready for injection(6).
|